Agilent Technologies N9340A manual Sweep Time, In non-zero spans, In zero span, Key Reference

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 86
Image 86

5 Key Reference

Sweep Time

 

Selects the length of time the analyzer takes to

 

tune across the displayed frequency span (or, in

 

zero span, the time the analyzer takes to sweep the

 

full screen) using the knob, the arrow keys, or

 

numeric keypad.

 

Reducing the sweep time increases the sweep rate.

N O T E

 

 

 

 

 

 

Key access: [BW/SWP]

 

In non-zero spans:

 

When the sweep time is auto- coupled, the analyzer

 

selects the optimum (shortest) sweep time ranging

 

from 10 ms to 1000 s for the current settings. This

 

selection process is influenced by three factors:

 

Frequency Span

 

The selected Resolution Bandwidth

 

The selected Video Bandwidth

 

Note that changing one of the three factors above

 

changes sweep time. The sweep time is in direct

 

ratio to frequency span and is in reverse ratio to

 

the selected resolution bandwidth and video

 

bandwidth when auto- coupled.

 

With sweep time set to a value less than the minimum

C ACUATUI O- N

 

auto-coupled value, an annotation “Data out of range”

 

appears in the active function block and the sweep time is

 

automatically converted to the minimum auto-coupled

 

value.

 

 

 

In zero span:

 

The auto- coupled status is invalid in zero span.

 

You can only set the sweep time manually. The

 

sweep time ranges from 6 ∝s to 200 s in zero span.