Agilent Technologies N9340A manual Press SPAN 50 MHz Press MARKER Peak Search, Making Measurements

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 48
Image 48

3Making Measurements

Press [MARKER] > {Peak Search}.

Press [MARKER] > {Marker To} > {To Center}.

4Change the span to 50 MHz and re- center the signal:

Press [SPAN] > 50 > {MHz}.

Press [MARKER] > {Peak Search}.

5Set the attenuation to 0 dB:

Press [AMPTD] > {Attenuation} > 0 > {dB}.

Press [MARKER] > {Marker To} > {To Ref}.

6To determine whether the harmonic distortion products are generated by the analyzer, first save the trace data in trace 2 as follows:

Press [TRACE] > {Trace (2)}.

Press [TRACE] > {Clear Write}.

7Allow trace 2 to update (minimum two sweeps), then store the data from trace 2 and place a delta marker on the harmonic of trace 2:

Press [TRACE] > {View}.

Press [MARKER] > {Peak Search}.

Press [Marker] > {Delta}.

Please refer to figure below shows the stored data in trace 2 and the measured data in trace 1. The Marker indicator reads the difference in amplitude between the reference and active markers.

38

N9340A User’s Guide

Page 48
Image 48
Agilent Technologies N9340A manual Press SPAN 50 MHz Press MARKER Peak Search, Press TRACE Trace Press TRACE Clear Write