Agilent Technologies N9340A manual Making a Stimulus Response Transmission Measurement

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 57
Image 57
Making a Stimulus Response Transmission Measurement

Making Measurements 3

Making a Stimulus Response Transmission Measurement

The procedure below describes how to use a built- in tracking generator to measure the rejection of a low pass filter, a type of transmission measurement.

1To measure the rejection of a low pass filter, connect the equipment as shown below.

A 370 MHz low- pass filter is used as a DUT in this example.

Figure 17 Transmission Measurement Test Setup

DUT

N9340A

1 2 3

4 5 6

7 8 9

0

2Press [Preset] to perform a factory preset.

3Set the start and stop frequencies and resolution bandwidth:

Press [FREQ] > {Start Freq} > 100 > {MHz}

Press [FREQ] > {Stop Freq} > 1 > {GHz}

Press [BW/SWP] > {RBW} > 1 > {MHz}

4Turn on the tracking generator and if necessary, set the output power to –10 dBm:

Press [Mode] > {Track Generator} > {Amplitude (On)} > –10 > {dBm}.

 

Excessive signal input may damage the DUT. Do not

C ACUATUI O- N

 

exceed the maximum power that the device under test can

 

tolerate.

 

 

N9340A User’s Guide

47

Page 57
Image 57
Agilent Technologies N9340A manual Making a Stimulus Response Transmission Measurement, Transmission Measurement Test Setup