| Getting Started 2 |
| Upgrading Firmware |
| When updating firmware, make sure there will be a |
C ACUATUI O- N | |
| constant power supply for at least 10 minutes. Power cut |
| off during the updating process can cause damage to the |
| instrument. |
A N9340A provides an easy and fast access for firmware update.
1Set up a folder named N9340DATA in the root directory of your USB memory stick.
2Store the firmware update application into the folder named N9340DATA.
3Insert the USB memory stick into the USB connector on the top panel of your analyzer.
4Press [SYS] > {More (2 of 3)} > {Update} > {Firmware} to activate the firmware updating procedure. Then the instrument will then perform the update automatically.
Adding an option
Pressing [SYS] > {More (2 of 3)} > {Option} > {Add} brings up a dialog box for entering the option
license code. Use the numeric keypad to input the option license code and use the [Enter] key as a
terminator. If the analyzer recognises the option license code, an message “Option install successfully” appears in the status line, otherwise a message “Invalid option licence” appears in the status line.
N9340A User’s Guide | 23 |