Agilent Technologies N9340A manual Control Limit line Buzzer, Control Limit Line Testing

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 114
Image 114

6 SCPI Command Reference

Control Limit line Buzzer

:CALCulate:LLINe[1]:BUZZer[:STATe]

OFFON01

:CALCulate:LLINe[1]:BUZZer[:STATe]?

This command toggles the audio warning between on and off. If data exceeds the limit line the buzzer will sound an alarm.

*RST state:

Off

Key access:

[Limit] > {Beep}

 

 

Control Limit Line Testing

:CALCulate:LLINe[1]:[STATe] OFFON01

:CALCulate:LLINe[1]:[STATe]?

This command toggles the limit line testing between on and off.

If the limit line is turned on, limit pattern is automatically turned off.

*RST state:

Off

Key access:

[Limit] > {Limit Line}

 

 

Set Limit Line Y-axis value

:CALCulate:LLINe[1]:Y <ampl>

:CALCulate:LLINe[1]:Y?

This command sets the Y- axis value of a limit line.

Limit line Y- axis value is set independently and is not affected by the X- axis units.

*RST value:

0 dBm

Available unit:

dBm, -dBm, mV, uV

Key access:

[Limit] > {Limit Line}

 

 

104

N9340A User’s Guide

Page 114
Image 114
Agilent Technologies N9340A manual Control Limit line Buzzer, Control Limit Line Testing, Set Limit Line Y-axis value