Agilent Technologies N9340A manual TRIGer Subsystem, External Trigger Slope, Trigger Source

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 149
Image 149

SCPI Command Reference 6

TRIGer Subsystem

Trigger subsystem is only valid when the analyzer is in zero span.

External Trigger Slope

:TRIGer[:SEQuence]:EXTernal:SLOPe

POSitiveNEGative

:TRIGer[:SEQuence]:EXTernal:SLOPe?

The external trigger signal must be a TTL signal at 0 V to 5 V. This function controls the trigger polarity (for positive or negative- going signals).

*RST state:

Positive

 

 

Trigger Source

:TRIGer[:SEQuence]:SOURce

IMMediateVIDeoEXTernal

:TRIGer[:SEQuence]:SOURce?

This command sets the source (or type) of triggering used to start a measurement.

Immediate is a free- run triggering.

Video triggers on the video signal level.

External allows you to connect an external trigger source.

*RST state:

Immediate (free- run triggering)

Key access:

[Span] > {Trigger}

 

 

N9340A User’s Guide

139

Page 149
Image 149
Agilent Technologies N9340A manual TRIGer Subsystem, External Trigger Slope, Trigger Source, SCPI Command Reference