Agilent Technologies N9340A manual Vbw/Rbw, Avg Type, Key Reference

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 84
Image 84

5 Key Reference

VBW

 

Changes the analyzer post- detector filter from

 

3 Hz to 1 MHz in a 1, 3, 10 sequence using the

 

knob, the step keys, or the numeric keypad.

 

If an unavailable bandwidth entered using the numeric

N O T E

keypad, the closest available bandwidth in the 1, 3, 10

 

 

sequence is used.

 

 

 

As the video bandwidth is decreased, the sweep

 

time is increased to maintain amplitude calibration.

 

A “#”mark appears next to RBW on the display

 

when it is not coupled.

 

Key access: [BW/SWP]

VBW/RBW

 

Selects the ratio between the video and resolution

 

bandwidths.

 

If signal responses near the noise level are visually

 

masked by the noise, the ratio can be set to less

 

than 1 to lower the noise. The knob and step keys

 

change the ratio in a 1, 3, 10 sequence.

 

When preset is set to default, or VBW/RBW is set to Auto,

N O T E

the ratio is set to 1.

 

 

 

 

Key access: [BW/SWP]

Avg Type

Accesses the functions to automatically or manually choose from one of the following averaging type:

74

N9340A User’s Guide

Page 84
Image 84
Agilent Technologies N9340A Vbw/Rbw, Avg Type, Key Reference, When preset is set to default, or VBW/RBW is set to Auto