SCPI Command Reference 6
DISPlay Subsystem
The DISPlay subsystem controls the selection and presentation of textual, graphical, and trace information. Within a display, information may be separated into individual windows.
Turn the Entire Display On/Off
:DISPlay:ENABle OFFON01
This command turns the display on or off. Having the display turned off may increase repetitive measurement rate.
*RST state: | On |
Trace Y-Axis Scaling
:DISPlay:WINDow:TRACe:Y[:SCALe] :PDIVision 12510
:DISPlay:WINDow:TRACe:Y[:SCALe]
:PDIVision?
This command sets the per- division display scaling
for the y- axis when scale type of Y axis is set to Log.
*RST value: | 10 dB |
Range: | 1 , 2 , 5 or 10 |
Key access: | [AMPTD] > {Scale/DIV} |
|
|
N9340A User’s Guide | 113 |