Agilent Technologies N9340A Resolving Signals of Equal Amplitude, Setup for obtaining two signals

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 38
Image 38
Resolving Signals of Equal Amplitude

3 Making Measurements

Resolving Signals of Equal Amplitude

In this example a decrease in resolution bandwidth is used in combination with a decrease in video bandwidth to resolve two signals of equal amplitude with a frequency separation of 100 kHz.

Notice that the final RBW selection to resolve the signals is the same width as the signal separation while the VBW is slightly narrower than the RBW.

1Connect two sources to the analyzer input as shown below.

Figure 3 Setup for obtaining two signals

N9310A

Figure 3 Setup for obtaining two signals•Press [PRESET].•Press [SPAN] > {Span} > 2 > {MHz}. 7 Manual background 8 Manual background 9Manual backgroundManual backgroundManual backgroundManual backgroundManual background 4 Manual background 5 Manual background 6Manual backgroundManual backgroundManual backgroundManual backgroundManual background

Manual backgroundManual background 1 Manual background 2 Manual background 3Manual backgroundManual backgroundManual background

Manual backgroundManual backgroundManual background 0 Manual background ·Manual backgroundManual backgroundManual background

Signal generator

Directional coupler

N9310A

 

 

7

8

9

4

5

6

1

2

3

0

·

 

Signal generator

2Set one source to 300 MHz. Set the frequency of the other source to 300.1 MHz. Set both source amplitudes to –20 dBm.

3Setup the analyzer to view the signals:

Press [PRESET].

Press [FREQ] > {Center Freq} > 300.05 > {MHz}.

Press [SPAN] > {Span} > 2 > {MHz}.

Press [BW/SWP] > {RBW} > 30 > {kHz}.

28

N9340A User’s Guide

Page 38
Image 38
Agilent Technologies N9340A Resolving Signals of Equal Amplitude, Setup for obtaining two signals, Making Measurements