Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

10.1.3.1

Signal I/O Diagnostic

 

 

 

Table 10-4: Diagnostic Mode - Signal I/O

 

 

 

 

 

 

 

No.

Signal

 

Control

Description

 

0.

DSP_BROWNOUT

 

NO

Display brownout is used to keep the display from

 

 

 

 

 

getting corrupted during low line voltage

 

 

 

 

 

conditions. Circuitry on the Status/Temp board

 

 

 

 

 

(01086) senses low line voltage and sets this bit.

 

 

 

 

 

The CPU reads this and generates the

 

 

 

 

 

BROWNOUT_RST signal described below.

 

 

 

 

 

 

 

1.

EXT_ZERO_CAL

 

NO

Shows state of status input bit to cause the

 

 

 

 

 

M100AH to enter Zero Calibration mode. Use to

 

 

 

 

 

check external contact closure circuitry.

 

 

 

 

 

 

 

2.

EXT_SPAN_CAL

 

NO

Shows state of status input bit to cause the

 

 

 

 

 

M100AH to enter the Span Calibration mode. Use

 

 

 

 

 

to check external contact closure circuitry.

 

3.

EXT_LOW_SPAN

 

YES

Shows state of status input bit to cause the

 

 

 

 

 

M100AH to enter the Low Span Calibration mode.

 

 

 

 

 

Use to check external contact closure circuitry.

 

4.

ZERO_VALVE

 

YES

Switches the Zero valve. Use this bit to test the

 

 

 

 

 

Zero valve function.

 

5.

CAL_VALVE

 

YES

Switches the Sample valve. Use this bit to test the

 

 

 

 

 

valve function.

 

 

 

 

 

 

 

6.

HIGH

 

YES

Energizes the high span valve. Use this bit to test

 

 

_SAPN_VALVE

 

 

the valve function.

 

 

 

 

 

 

 

7.

RCELL_ HEATER

 

NO

Shows the status of the reaction cell heater. This has

 

 

 

 

 

the same function as the LED in the power supply

 

 

 

 

 

module.

 

 

 

 

 

 

 

8.

ELEC_TEST

 

YES

Turns on electric test bit in Preamp. Should be used

 

 

 

 

 

for troubleshooting Preamp circuit. We recommend

 

 

 

 

 

you use the ELEC TEST button in the DIAG menu

 

 

 

 

 

to operate electric test.

 

 

 

 

 

 

 

9.

OPTIC_TEST

 

YES

Turns on optic test bit in Preamp. Should be used

 

 

 

 

 

for isolating PMT detector system from the rest of

 

 

 

 

 

the Preamp circuit. We recommend you use the

 

 

 

 

 

OPTIC TEST button in the DIAG menu to operate

 

 

 

 

 

optic test.

 

 

 

 

 

 

(table continued)

10-13

Page 121
Image 121
Teledyne 100AH instruction manual Signal I/O Diagnostic Diagnostic Mode Signal I/O, 10-13