Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

Table 10-3: Summary of Diagnostic Modes

DIAG Mode

Description

SIGNAL I/O

Gives access to the digital and analog inputs and outputs on the V/F

 

board. The status or value of all of the signals can be seen. Some of

 

the signals can be controlled from the keyboard. Table 10-4gives

 

details on each signal and information on control capabilities.

 

NOTE: Some signals can be toggled into states that indicate

 

warnings or other faults. These settings will remain in effect until

 

DIAG mode is exited, then the M100AH will resume control over

 

the signals.

 

 

ANALOG OUTPUT

Causes a test signal to be written to the analog output DAC's. The

 

signal consists of a scrolling 0%, 20%, 40%, 60%, 80%, 100% of

 

the analog output value. The scrolling may be stopped by pressing

 

the key underneath the % display to hold that value. The exact

 

voltage values depend on the jumper settings on the analog output

 

buffer amplifiers.

 

 

DAC CALIBRATION

The analog output is created by 4 digital-to-analog converters. This

 

selection starts a procedure to calibrate these outputs. Refer to

 

Section 10.3.2.1 for a detailed procedure.

OPTICAL TEST

Sets the M100AH into a known state and turns on an LED near the

 

PMT to test the instrument signal path. See Section 10.1.3.3 for

 

details on using this test.

ELECTRICAL TEST

Tests just the electronic portion of the PMT signal path. Used in

 

conjunction with optic test, see Section 10.1.3.2.

LAMP CAIBRATION

This feature allows to update the Lamp Calibration value. Displayed

 

value is the current lamp intensity and pressing ENTR key will

 

update the Lamp Calibration value. Refer Section 10.4.2 for UV

 

Lamp adjustment. See also Section 10.1.6 Factory Calibration

 

Procedure.

 

 

FLOW CALIBRATION

This feature allows to enter the actual sample flow. Once the flow is

 

calibrated, it will compute the flow automatically from the pressure

 

measurements. Refer Section 10.3.5.

 

 

TEST CHANNEL

This feature allows to output scaled voltage of most test

OUTPUT

measurement through the analog output terminal. Refer

 

Section 10.1.5.

 

 

RS-232

Causes a 1 second burst of data to be transmitted from the RS-232

 

port. Used to diagnose RS-232 port problems. See Sections 10.1.3.6

 

and 4.3.1.2 for RS-232 port diagnostic techniques.

 

 