Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

Table 10-1: Test Functions (Continued)

Test Function

Factory Set-Up

Comment

UV LAMP

3000 - 4000 mV

This is the instantaneous reading of the UV lamp intensity.

 

 

Typical UV lamp intensity is between 2000 mV and

 

 

4000 mV. See Section 10.4.2 for how to peak the UV lamp

 

 

output. Intensity lower than 600 mV will cause WARNING

 

 

and below 350 mV will cause to display XXXXX.X.

 

 

Low UV lamp intensity could be due to:

 

 

1. Aging UV lamp.

 

 

2.

Beam splitter out of alignment.

 

 

3.

Faulty lamp transformer.

 

 

4.

Aging or faulty UV detector.

 

 

5.

Dirty optical components.

 

 

 

LAMP RATIO

80 - 120%

This is the computed value which is the ratio of current UV

 

 

LAMP reading respect to the calibrated UV LAMP reading.

 

 

Usually the ratio value drops as the aging lamp output decays.

 

 

Low LAMP RATIO can cause the slope and offset out of

 

 

typical range.

 

 

See Section 10.4.2 how to reset the lamp calibration and

 

 

LAMP RATIO.

STR LGT

<50 PPM

Stray light is the background light of the reaction cell

 

 

expressed in PPM while sampling zero gas.

 

 

High stray light could be caused by :

 

 

1.

Aging UV filter.

 

 

2.

Contaminated reaction cell.

 

 

3.

Light leak.

 

 

4.

Pneumatic leak.

 

 

 

DRK PMT

-50 ± 200 mV

This is the reading of the PMT signal without the UV lamp.

 

 

When the lamp is momentarily turned off, the darkness inside

 

 

of the reaction cell is monitored and compensated for any UV

 

 

detector dark current drift or offset.

 

 

High dark PMT reading could be due to:

 

 

1.

Light leak.

 

 

2.

Lamp not turning off completely.

 

 

3. High PMT temperature.

 

 

4.

High electronic offset.

 

 

 

 

(table continued)

10-6

Page 114
Image 114
Teledyne 100AH instruction manual 10-6, Lamp Ratio, Str Lgt, Drk Pmt