Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

10.1.2 Fault Diagnosis with WARNING Messages

The M100AH monitors several internal values for alarm conditions. If the condition for an alarm is met, the alarm is displayed on the front panel and the warning is transmitted out the RS-232 port. Any time the instrument is powered up the SYSTEM RESET alarm will be displayed. Generally, it is ok to ignore warnings that are displayed shortly after power-up; only if they persist should they be investigated.

Table 10-2shows the warning messages and gives some possible causes.

Table 10-2: Front Panel Warning Messages

Message

Description

SYSTEM RESET

Analyzer was reset/powered on. This warning occurs every time the

 

instrument is powered up, as in a power failure. It can also occur if

 

the RAM or EEPROM is reset.

 

 

RAM INITIALIZED

RAM was erased. The RAM contains the DAS averages which get

 

erased when the RAM is initialized. It also contains temporary data

 

used by the M100AH to calculate concentrations. No setup

 

variables are stored in the RAM.

 

 

PMT DET WARNING

The PMT output above 4995 mV. The Preamp circuit could be out

 

of adjustment or wrong range selected.

 

 

UV LAMP WARNING

The UV lamp output is below 600 Mv or above 4995 Mv. UV lamp

 

could be aging or need to peak output by adjusting the position.

 

Also beam splitter out of alignment will cause to read lower.

DARK CAL WARNING

Dark PMT or UV lamp reading is above 400 mV.

 

 

HVPS WARNING

The HVPS control voltage is above 900 V or below 400 V. The

 

Preamp circuit could be out of adjustment or the PMT sensitivity is

 

too weak.

SAMPLE FLOW

The measured sample flow is outside the hi/low limits. A Leak in

WARNING

the pneumatic system is the main cause of the warning although the

 

flow sensor itself could be the cause.

 

 

SAMPLE PRESS

The sample pressure is below 15”Hg or above 35”Hg.

WARNING

 

 

 

(table continued)

10-9

Page 117
Image 117
Teledyne 100AH Fault Diagnosis with Warning Messages, Front Panel Warning Messages, 10-9, Sample Flow, Sample Press