Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

Table 2-1: Final Test and Calibration Values (Continued)

Parameter

Observed

Units

Nominal Range

Reference Section

Value

 

 

 

 

 

 

 

 

 

Span and Cal Values

 

SO2 Span Conc

 

PPM

100 - 4500

 

Table 8-3

SO2 Slope

 

 

1.0 ± 0.3

 

8.7

 

 

 

 

 

 

 

SO2 Offset

 

mV

<100

 

8.7

Noise at Zero

 

PPM

0.05 - 0.2

 

Table 10-1

(rms)

 

 

 

 

 

 

 

 

 

 

 

 

Noise at Span

 

PPM

0.5% of reading

Table 10-1

(rms)

 

 

(above 50PPM)

 

 

 

 

 

 

 

 

 

 

 

Measured Flows

 

Sample Flow

 

cc/min

650 ± 60

 

10.2.2, 10.3.5,

 

 

 

 

 

 

Figure 10-5

Factory Installed Options

 

Option Installed

 

Power Voltage/Frequency

 

 

 

 

 

 

 

 

 

 

Rack Mount, w/ Slides

 

 

 

 

 

 

 

 

 

 

Rack Mount, w/ Ears Only

 

 

 

 

 

 

 

 

 

 

Zero/Span Valves Manifold

 

 

 

 

 

 

 

 

 

 

 

Multi-drop

 

 

 

 

 

 

 

 

 

 

 

 

Kicker

 

 

 

 

 

 

 

 

 

 

 

 

37 mm Filter

 

 

 

 

 

 

 

 

 

 

4-20 mA Isolated Current Loop Output

ISOLATED

NON-ISOLATED

(non-isolated standard)

 

 

 

 

 

SO2 (RANGE 1)

 

 

____________

____________

SO2 (RANGE 2)

 

 

____________

____________

SPARE

 

 

____________

____________

TEST OUTPUT

 

 

____________

____________

 

 

 

 

 

 

 

PROM #

 

 

 

Serial #

 

 

Date

 

 

 

Technician

 

 

2-13