Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

The next task is to calibrate the analyzer. There are several ways to do a calibration; they are summarized in Table 7-1.For a preliminary checkout we recommend calibration with zero air and span gas coming in through the sample port. The procedure is:

WARNING – Analyzer Exhaust

Danger – Analyzer exhaust may contain high concentration of SO2 gas. Exhaust properly from the pump pack exhaust to well ventilated area at atmospheric pressure.

Step 1 - Enter the expected SO2 span gas concentration:

Step Number

Action

Comment

1.

Press CAL-CONC

This key sequence causes the M100AH to prompt for the

 

 

expected SO2 concentration. Enter the SO2 span

 

 

concentration value by pressing the key under each digit until

 

 

the expected value is set.

 

 

 

2.

Press ENTR

ENTR stores the expected SO2 span value. This value will be

 

 

used in the internal formulas to compute subsequent SO2

 

 

concentration values.

3.

Press EXIT

Returns instrument to SAMPLE mode.

 

 

 

4.

Press SETUP-

If necessary you may want to change ranges. Normally the

 

RNGE-SET

instrument is shipped in single range mode set at 500 PPM.

 

 

(see Section 5.3.4 for Range Menu)

 

 

 

5.

Press EXIT

Returns the instrument to SAMPLE mode.

 

 

 

2-8

Page 20
Image 20