Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

3SPECIFICATIONS, AGENCY APPROVALS, WARRANTY

3.1 Specifications

Ranges

In 1 PPM increments from 10 PPM to 5000 PPM, dual ranges or autoranging

Noise at Zero

0.05 PPM RMS

Noise at Span

<0.5% of reading (above 50 PPM)

Lower Detectable Limit1

0.1 PPM RMS

Zero Drift2

< 1 PPM/24 hours

Zero Drift2

<2 PPM/7 days

Span Drift2

<0.5% FS/7 days

Lag Time

5 sec

Rise Time

95% in < 30 sec

Fall Time

95% in < 30 sec

Sample Flow Rate

650cc/min. ±10%

Linearity

1% of full scale

Precision

0.5% of reading

Temperature Range

5 - 40oC

Temp Coefficient

< 0.1% per oC

Humidity

0 - 95% RH, non-condensing

Voltage Coefficient

< 0.05% per V

Dimensions HxWxD

7"x17"x23.6" (18 cm x 43 cm x 61 cm)

Weight, Analyzer

43 lbs (19.5 kg)

Weight, Pump Pack

16 lbs (7 kg)

Power, Analyzer

110 v∼60 Hz, 220 v∼50 Hz, 240 v∼50 Hz, 250 watts

Power, Analyzer3

230 v∼50 Hz, 2.5A

Power, Ext. Pump

110 v∼60 Hz, 220 v∼50 Hz, 240 v∼50 Hz, 295 watts

Power, Ext. Pump3

230 v∼50 Hz, 2.5 A

Environmental

Installation Category (Overvoltage Category) II

Conditions

Pollution Degree 2

Recorder Output4

4 - 20 mA non-isolated standard, 0-100 mV, 0-1, 5, 10 v ; resolution of 1 part

 

in 1024 of selected voltage or current range. 4 - 20 mA isolated optional.

Status Option

12 Status Outputs from opto-isolator

Measurement Units

PPM, mg/m3

1.Defined as twice the zero noise level.

2.At constant temperature and voltage.

3.Electrical ratings for CE Mark compliance.

4.Bi-polar. (voltage or current selectable by the jumper on the motherboard)

3-1

Page 27
Image 27
Teledyne 100AH instruction manual SPECIFICATIONS, Agency APPROVALS, Warranty, Specifications, Ppm Rms