Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

10.3.4 Power Supply Module

The Power Supply Module consists of several subassemblies described in Table 10-7below.

Table 10-7: Power Supply Module Subassemblies

Module

Description

Linear Power Supply

The linear power supply board takes multiple voltage inputs from

Board

the power transformer and produces +5, +15, -15, +12 VDC

 

outputs. The outputs are routed to two external connectors, P2 and

 

P3. See Figure 10-4.The +5 is used for operating the CPU. The

 

± 15 is used in several locations for running op-amps and IC's. The

 

+12 is used for operating fans and valves.

Switching Power Supply

The switching power supply supplies +15 VDC at 4 A to the PMT

 

cooler control on the Sensor Module. The output is made available

 

through J10 on the Switch Board. There is a load resistor on the

 

Switch Board to keep the output stable when little current is

 

required from the supply.

 

 

Switch Board

The Switch Board has many different functions. It takes logic

 

signals from the V/F board and uses them to switch 4-115 VAC and

 

4-12 VDC loads. The board also contains the instrument central

 

grounding tie point. It distributes AC and DC power as needed.

 

Connector J2 programs the power transformers to take 115, 220, or

 

240 VAC inputs

 

 

Power Transformers

There are potentially 2 input power transformers in the instrument.

 

The multitap transformer T1 is in every M100AH and supplies input

 

power for the Linear Power Supply board described above. A

 

second transformer T2 is added if 220 or 240 VAC input is required.

 

Input power selection is done via a programming connector P2

 

which provides the proper connections for either foreign or

 

domestic power.

 

 

Circuit Breaker/Power

The front panel contains a combination circuit breaker - input power

Switch

switch. It is connected to the PSM through J6 on the Switch Board.

 

If an overload is detected the switch goes to the OFF position.

 

Switching the power back on resets the breaker also.

 

 

10-41