Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D
NOTE
Avoid setting two or more sequences at the same time of the day. Any new
sequence which is initiated whether from a timer, the RS-232, or the contact
closure inputs will override any sequence that is in progress.
The programmed start time must be a minimum of 5 minutes
later than the real time clock.
Examples of possible sequences are as following under any one of three available SEQx.
Example 1: To perform zero-span calibration check once per day at 10:30 PM, 5/20/97.
1) MODE: ZERO-HI
2) TIMER ENABLE: ON
3) STARTING DATE: 5/20/97
4) STARTING TIME: 22:30
5) DELTA DAYS: 1
6) DELTA TIME: 00:00
7) DURATION: 15.0 MINUTES
8) CALIBRATE: OFF
Example 2: To perform zero calibration adjust once per day retarding 15 minutes everyday starting
at 11:30 pm, 5/20/97.
1) MODE: ZERO
2) TIMER ENABLE: ON
3) STARTING DATE: 5/20/97
4) STARTING TIME: 23:30
5) DELTA DAYS: 0
6) DELTA TIME: 23:45
7) DURATION: 15.0 MINUTES
8) CALIBRATE: ON
6-4