Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

NOTE

Avoid setting two or more sequences at the same time of the day. Any new sequence which is initiated whether from a timer, the RS-232, or the contact closure inputs will override any sequence that is in progress.

The programmed start time must be a minimum of 5 minutes

later than the real time clock.

Examples of possible sequences are as following under any one of three available SEQx.

Example 1: To perform zero-span calibration check once per day at 10:30 PM, 5/20/97.

1)MODE: ZERO-HI

2)TIMER ENABLE: ON

3)STARTING DATE: 5/20/97

4)STARTING TIME: 22:30

5)DELTA DAYS: 1

6)DELTA TIME: 00:00

7)DURATION: 15.0 MINUTES

8)CALIBRATE: OFF

Example 2: To perform zero calibration adjust once per day retarding 15 minutes everyday starting at 11:30 pm, 5/20/97.

1)MODE: ZERO

2)TIMER ENABLE: ON

3)STARTING DATE: 5/20/97

4)STARTING TIME: 23:30

5)DELTA DAYS: 0

6)DELTA TIME: 23:45

7)DURATION: 15.0 MINUTES

8)CALIBRATE: ON

6-4