Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

10.1.6.1 Electric Test (ET) Procedure

1.Any time the gain of the PMT buffer amplifier circuit (R19) is changed (refer component location label inside of the Preamp cover), that will change the ET reading. Since the gain was adjusted in the above procedure, do the following to re-adjust the ET signal.

2.To re-adjust ET press SETUP-MORE-DIAG, then scroll to ELEC TEST and press ENTR.

3.Scroll the TEST functions until PMT is displayed.

4.Adjust R27 until 2000 mV ± 50 is displayed.

5.Press EXIT to return to SAMPLE mode.

10.1.6.2 Optic Test (OT) Procedure

1.Any time the PMT buffer amplifier circuit (R19) gain and/or the PMT high voltage setting is changed, the OT reading will also change. The PMT cal procedure changed both of these values.

2.To re-adjust OT press SETUP-MORE-DIAG, then scroll to OPTIC TEST and press ENTR.

3.Scroll the TEST functions until PMT is displayed.

4.Adjust R25 until 200 mV ± 20 is displayed.

5.Press EXIT to return to SAMPLE mode.

10-22

Page 130
Image 130