Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

8.7 Calibration Quality

After Zero/Span is complete, it is very important to check the QUALITY of the calibration. The calibration of the M100AH involves balancing several sections of electronics and software to achieve an optimum balance of accuracy, noise, linearity and dynamic range.

The following procedure compares the Slope and Offset parameters in the equation used to compute the SO2 concentration.

The slope and offset parameters are similar to the span and zero pots on an analog instrument. Just as in the analog instrument, if the slope or offset gets outside of a certain range, the instrument will not perform as well.

The offset value gives information about the background signal level. Check the observed offset value against the factory value in Table 2-1.If significantly higher check Section 10.1.6. Increasing readings are a predictor of problems.

Table 8-8: Calibration Quality Check

Step

Action

Comment

Number

 

 

1.

Scroll the TEST

Typical SLOPE value for SO2 is 1.0 ± 0.3. If the value is

 

function menu until the

not in this range, check Section 10.1.6. If the SLOPE value

 

SLOPE is displayed.

is in the acceptable range the instrument will perform

 

 

optimally.

 

 

 

2.

Scroll the TEST

Typical number is less than 200mV which is mainly the

 

function menu until the

optical system background. If the OFFSET value is outside

 

OFFSET is displayed.

this range, check Section 10.1.6.

 

 

 

After the above procedure is complete, the M100AH is ready to measure sample gas.

8-8

Page 96
Image 96
Teledyne 100AH instruction manual Calibration Quality Check, Step Action Comment