Table 10-1: Test Functions (Continued)

Test Function

Factory Set-Up

Comment

PMT

0 - 5000 mV

This is the instantaneous output of the PMT.

 

 

The PMT voltage values will be relatively constant when:

 

 

1.

Electric test - variation in the 2000 mV signal observed will

 

 

 

be noise of the V/F board and preamp noise. See Section

 

 

 

10.1.3.2.

 

 

2.

Optic test - variation in the 400 mV signal will be PMT

 

 

 

dark current, preamp, HVPS, and electronic. See Section

 

 

 

10.1.3.3.

 

 

3.

Sampling zero gas.

 

 

4.

Sampling stable SO2 span gas.

When sampling zero gas the PMT reading should be less than 50 mV and relatively constant.

High or noisy readings could be due to:

1.Pneumatic leak.

2.Excessive background light which is caused by a possible aging UV filter.

3.Low UV lamp output.

4.PMT recently exposed to room light. It takes 24-48 hours for the PMT to adapt to dim light.

5.Light leak in reaction cell.

6.Reaction cell contaminated.

7.Vacuum pressure reading not stable.

8.HVPS not stable.

(table continued)

10-5

Page 113
Image 113