Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

Table 10-4: Diagnostic Mode - Signal I/O (Continued)

No.

Signal

Control

Description

31.

HVPS_VOLTAG

NO

HVPS programming voltage. Output of HVPS is

 

 

 

1000x value present.

 

 

 

 

32.

DAC_CHAN_0

NO

Output of SO2 (DAC0/RANGE1) in mV.

33.

DAC_CHAN_1

NO

Output of SO2 (DAC1/RANGE2) in mV.

34.

DAC_CHAN_2

NO

Output of spare (DAC2) channel in mV.

 

 

 

 

35.

DAC_CHAN_3

NO

Test Channel (DAC3) output.

 

 

 

 

36.

UVLAMP_SIGNAL

NO

Current UV lamp signal voltage in mV. Same as

 

 

 

UV LAMP in TEST menu. Bi-polar, typically 0-

 

 

 

5000 mV range. A constant value of 5000 mV

 

 

 

indicates off scale.

 

 

 

 

37.

CONC_OUT_1

YES

DAC0 (SO2/RANGE1) analog output in mV.

 

 

 

 

38.

CONC_OUT_2

YES

DAC1 (SO2/RANGE2) analog output in mV.

 

 

 

 

39.

TEST_OUTPUT

YES

DAC3 (TEST CHANNEL) analog output in mV.

 

 

 

 

10.1.3.2 Electric Test

This function injects a constant voltage between the preamplifier and the buffer amplifier on the preamp board. Electric test checks part of the Preamp, the V/F and computer for proper functioning. The result of electric test should be a smooth quiet signal as shown by constant values for the SO2 concentration. Likewise the analog outputs should produce a smooth quiet trace on a strip chart (analog output range is set to 5000PPM and auto-ranging is disabled).

Procedure:

1.Scroll the TEST function to PMT.

2.Press SETUP-MORE-DIAG, scroll to ELECT TEST by pressing the NEXT button. When ET appears, press ENTR to turn it on.

3.The value in PMT should come up to 2000 mV ± 100 mV in less than 15 sec.

10-16

Page 124
Image 124
Teledyne 100AH instruction manual Electric Test, 10-16