Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

6 OPTIONAL HARDWARE AND SOFTWARE

Optional equipment offered with the M100AH includes:

1.Rack mount with slides (P/N 00280)

2.Rack mount without slides, ears only (P/N 01470)

3.Zero/Span valves assembly (P/N 0178603)

4.4-20mA, isolated outputs (P/N 01471)

6.1Rack Mount Options

The Rack Mount option including slides and ears, permits the Analyzer to be mounted in a standard 19" wide x 24" deep RETMA rack. The Rack Mount option can also be ordered without slides for applications requiring the instrument to be rigidly mounted in a RETMA rack.

6.2 Zero/Span Valves

The Zero/Span Valve option consists of a manifold with four valves. See Figure 2-5for valve location. Connections are provided on the rear panel for two (low and high) span gas and zero gas inputs to the valves (See Figure 2-2). These valves can be actuated by several methods shown in Table 6-1.

Table 6-1: Zero/Span Valve Operation

Mode

Description

Reference Section

1.

Front panel operation via

Calibration Section 8 - Manual Zero/Span Check.

 

CALS and CALZ

 

 

buttons.

 

 

 

 

2.

Automatic operation

Setup and use of AUTOCAL is described in Table 6-2,

 

using AUTOCAL

and Section 8.3.

 

 

 

3.

Remote operation using

Setup described in Table 7-1.Operation of AUTOCAL

 

the RS-232 interface

described in Section 6.3 and Section 7.5.

 

 

 

4.

Remote operation using

Section 8.5 - Automatic operation using external contact

 

external contact closures

closures. Table 8-7.

 

 

 

6-1

Page 65
Image 65
Teledyne 100AH Optional Hardware and Software, Rack Mount Options, Zero/Span Valves, Zero/Span Valve Operation