Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

SETUP PROPERTIES FOR CONC:

NAME:

CONC

EVENT:

ATIMER

STARTING DATE:

25-JUL-96

SAMPLE PERIOD:

000:00:01

REPORT PERIOD:

000:00:05

NUMBER OF RECORDS:

800

RS-232 REPORT:

ON

COMPACT REPORT:

OFF

CHANNEL ENABLED:

ON

CAL. HOLD OFF:

ON

PARAMETERS:

1

PARAMETER=CONC1, MODE=AVG, PRECISION=1

In this example, the data channel’s NAME property is “CONC”; the EVENT property is ATIMER; the PARAMETERS property is 1 (indicating a single parameter); the NUMBER OF RECORDS property is 800, and the RS-232 REPORT property is ON. The list of parameters and their properties is also printed. Each data channel stores its data in a separate file in the RAM disk, and this property shows the file name.

Data Report Format

A data report format looks like the following:

D 31:10:06 0412 CONC : AVG CONC1=6.8 PPM

This report uses the traditional TELEDYNE API format of a leading first character (“D” in this example), a time stamp (“31:10:06”), and the instrument ID (“0412”). The other fields in the report are the data channel name (“CONC”), the sampling mode (“AVG”), the parameter (“CONC1”), the parameter value (“6.8”), and the units (“PPM”).

If the RS-232 interface is in the quiet mode, then these reports are not printed, although they can be requested by a user or host computer at a later time.