Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

Table 10-5: Model 100AH Variables

No.

Name

Units

Default

Value

Description

Value

Range

 

 

 

 

0.

DAS_HOLD_

MIN

15.0

0.5-20.0

Hold off duration after

 

OFF

 

 

 

calibration or diagnostic

 

 

 

 

 

mode.

 

 

 

 

 

 

1.

TPC_ENABLE

 

ON

ON-OFF

Temp/Pres compensation

 

 

 

 

 

enable

 

 

 

 

 

 

2.

RCELL_SET

°C

50

30-70

Reaction cell temperature

 

 

 

 

 

set point

3.

DYN_ZERO

 

OFF

OFF/ON

Enable to adjust zero

 

 

 

 

 

calibration through remote

 

 

 

 

 

contact closure

 

 

 

 

 

 

4.

DYN_SPAN

 

OFF

OFF/ON

Enable to adjust span

 

 

 

 

 

calibration through remote

 

 

 

 

 

contact closure

 

 

 

 

 

 

5.

RS232_MODE

Bit

8

0-99999

Value is SUM of following

 

 

Field

 

 

decimal numbers:

 

 

 

 

 

1=enable quiet mode

 

 

 

 

 

2=enable computer mode

 

 

 

 

 

4=enable security feature

 

 

 

 

 

8=enable front panel RS-

 

 

 

 

 

232 menus (Teledyne API

 

 

 

 

 

protocol)

 

 

 

 

 

16=enable alternate

 

 

 

 

 

protocol (i.e. Hessen)

 

 

 

 

 

32=enables multi-drop

 

 

 

 

 

support

6.

CLOCK_ADJ

Sec.

0

±60

Real-time clock adjustment

 

 

 

 

 

 

10-19

Page 127
Image 127
Teledyne instruction manual Model 100AH Variables, 10-19, Name Units Default Value Description Range