Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

Table 10-1: Test Functions (Continued)

Test Function

Factory Set-Up

Comment

DRK LMP

-50 ± 200 mV

This is the reading of the UV reference detector without the

 

 

UV lamp. When the is momentarily turned off, UV detector

 

 

signal is the dark current of the detector itself. The dark

 

 

current is monitored and compensated for any UV detector

 

 

dark current drift or offset.

 

 

High dark UV detector could be caused by:

 

 

1.

Lamp not turning off completely.

 

 

2.

High electronic offset.

 

 

 

SLOPE

1.0 ± 0.2

The slope can be thought as a gain term which determines the

 

 

steepness of the calibration curve.

 

 

M100AH will operate normally even the slope is out of range,

 

 

however it is preferred within the range for optimum

 

 

operation of the analyzer.

 

 

Slope out of range could be due to:

 

 

1.

Analog gain pot needs adjustment (see Section 10.1.6).

 

 

2.

Poor calibration quality (see Section 8.7).

 

 

 

OFFSET

<50 mV

This is essentially identical to the stray light except it is

 

 

expressed in mV.

 

 

High offset could be due to:

 

 

1.

Light leak.

 

 

2.

Aging UV filter.

 

 

3.

Contaminated reaction cell.

 

 

4.

Pneumatic leak.

 

 

5.

Poor calibration quality.

 

 

 

HVPS

400 - 900 V

This represents the scaled-up HVPS programming voltage to

 

 

the HVPS. The design of the HVPS precludes taking a single

 

 

reading that indicates the health of the supply. Refer to the

 

 

HVPS Troubleshooting Section 10.3.9 for a procedure for

 

 

testing the HVPS. This TEST function is used primarily to set

 

 

the HVPS voltage value and the reading should be very stable

 

 

and constant. A value not in the 400 to 900 volt range

 

 

indicates problems with the HVPS supply.

 

 

 

 

(table continued)

10-7

Page 115
Image 115
Teledyne 100AH instruction manual 10-7, Drk Lmp