Teledyne 100AH Diagnostic Commands and Messages, 11 RS-232 Diagnostic Command Summary

Models: 100AH

1 172
Download 172 pages 44.73 Kb
Page 83
Image 83

Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

7.6 DIAGNOSTIC Commands and Messages

When Diagnostic mode is entered from the RS-232 port, the diagnostic mode issues additional status messages to indicate which diagnostic test is currently selected. Examples of Diagnostic mode messages are:

D DDD:HH:MM IIII ZERO VALVE=ON

D DDD:HH:MM IIII ENTER DIAGNOSTIC MODE

D DDD:HH:MM IIII EXIT DIAGNOSTIC MODE

The following is a summary of the Diagnostic commands.

Table 7-11: RS-232 Diagnostic Command Summary

Command

Description

D LIST

Prints all I/O signal values. See Table 10-4for Signal I/O

 

definitions.

 

 

D name=value

Examines or sets I/O signal. For a list of signal names see

 

Table 10-4in Section 10. Must issue D ENTER SIG command

 

before using this command.

 

 

D LIST NAMES

Prints names of all diagnostic tests.

 

 

D ENTER SIG

Executes SIGNAL I/O diagnostic test.

D ENTER OT

Executes Optic Test diagnostic test.

D ENTER ET

Executes Elect Test diagnostic test.

D ENTER TASK

Displays a listing of the tasks and their status.

 

Use D EXIT to leave these diagnostic modes.

 

 

D EXIT

Must use this command to exit SIG, ET or OT Diagnostic modes.

 

 

D RESET

Resets analyzer software (same as power on).

 

 

D RESET RAM

Resets analyzer software and erases RAM. Erases SO2

 

concentration values. Keeps setup variables and calibration.

 

(same as installing new software version)

 

 

D RESET EEPROM

Resets analyzer software and erases RAM and EEPROM. Returns

 

all setup variables to factory defaults, resets calibration value.

7-13

Page 83
Image 83
Teledyne 100AH instruction manual Diagnostic Commands and Messages, 11 RS-232 Diagnostic Command Summary