Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

4.3.2 Status Output Interface

The status output is a feature that reports the Analyzer conditions via contact closures on the rear panel. The closures are available on a 50 pin connector on the rear panel. The contacts are NPN transistors which can draw up to 50 mA of DC current. The pin assignments are listed in the Table below.

Table 4-3: Status Output Pin Assignments

Output #

PIN #

Definition

Condition

1

1

(-), 2 (+)

ZERO CAL

CLOSED IN ZERO CAL

 

 

 

 

 

2

3

(-), 4 (+)

SPAN CAL

CLOSED IN SPAN CAL

 

 

 

 

 

3

5

(-), 6 (+)

FLOW ALARM

CLOSED IF FLOW WARNING

 

 

 

 

 

4

7

(-), 8 (+)

TEMP ALARM

CLOSED IF ANY TEMP WARNING

 

 

 

 

 

5

9

(-), 10 (+)

DIAG MODE

CLOSED IN DIAG MODE

 

 

 

 

6

11 (-), 12 (+)

POWER OK

CLOSED IF SYSTEM POWER OK

 

 

 

 

7

21 (-), 22 (+)

SYSTEM OK

CLOSED IF SYSTEM OK

 

 

 

 

8

19 (-), 20 (+)

HVPS ALARM

CLOSED IF HVPS WARNING

 

 

 

 

9

13 (-), 14 (+)

SPARE

 

 

 

 

 

10

23 (-), 24 (+)

HIGH RANGE

CLOSED IF HIGH PMT RANGE

 

 

 

 

11

25 (-), 26 (+)

LOW SPAN CAL

LOW SPAN CALIBRATION

 

 

 

 

12

27 (-), 28 (+)

UV LAMP ALARM

CLOSED IF UV LAMP WARNING

 

 

 

 

 

The Status/Temp Board schematic can be found in the Appendix Drawing 01087.

4-11

Page 41
Image 41
Teledyne 100AH instruction manual Status Output Interface, Status Output Pin Assignments, Output #, Definition Condition