Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

Table 7-5: RS-232 Command Summary

Commands

Definition

?

Print help screen

 

 

T SET ALL

Enable display of all test variables during T LIST

 

 

T SET name

Display only NAME during T LIST

 

 

T LIST [ALL]

Print all test variables enabled with T SET or ALL

 

warnings

 

 

T name

Print single test, "name" from Table 7-7

 

 

T CLEAR ALL

Disable T LIST, use with T SET name

 

 

W SET ALL

Enable display of all warnings during W LIST

 

 

W LIST [ALL]

Print warnings enabled with W SET or ALL warnings

 

 

W name

Print individual "name" warning from Table 7-8

 

 

W CLEAR ALL

Disable W LIST, use with W SET

 

 

C command

Execute calibration "command" from Table 7-10

 

 

D LIST

Prints all I/O signal values

 

 

D name

Prints single I/O signal value/state

 

 

D name=value

Sets variable to new "value"

 

 

D LIST NAMES

Lists diagnostic test names

 

 

D ENTER name

Enters and starts 'name' diagnostic test

 

 

D EXIT

Exits diagnostic mode

 

 

D RESET

Resets analyzer (same as power-on)

 

 

D RESET RAM

System reset, plus erases RAM. Initializes DAS, SO2

 

concentration readings, calibration not affected.

 

 

D RESET EEPROM

System reset, plus erases EEPROM (RAM_RESET

 

actions + setup variables, calibration to default values)

 

 

V LIST

Print all easy variable names from Table 10-5

 

 

V name

Print individual "name" variable

 

 

V name=value

Sets variable to new "value"

 

 

V CONFIG

Print analyzer configuration

 

 

7-6

Page 76
Image 76
Teledyne 100AH instruction manual RS-232 Command Summary, Commands Definition