Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

5.3Set-Up Mode

5.3.1Configuration Information (CFG)

This menu item will tell if the installed software has factory special features or other non- standard features. If you call Teledyne API service, you may be asked for information from this menu.

5.3.2 Automatic Calibration (AutoCal)

The AutoCal feature allows the M100AH to automatically operate the Zero/Span Valve option ON on a timed basis to check or adjust its calibration. This menu item is shown only if the Zero/Span Valve option is installed. Detailed information on setting up AutoCal is found in the Section 6.3.

5.3.3 Data Acquisition System (DAS)

The Model 100AH contains a flexible and powerful built in data acquisition system (DAS) that enables the analyzer to store concentration data as well as many diagnostic parameters in its battery backed memory. This information can be viewed from the front panel or printed out through the RS-232 port. The diagnostic data can be used for performing “Predictive Diagnostics” and trending to determine when maintenance and servicing will be required.

The logged parameters are stored in what are called “Data Channels.” Each Data Channel can store multiple data parameters. The Data Channels can be programmed and customized from the front panel. A set of default Data Channels has been included in the Model 100AH software. These are described Section 5.3.3.1. For more information on programming custom Data Channels, a supplementary document containing this information can be requested from Teledyne API.

5.3.3.1 Data Channels

The function of the Data Channels is to store, report, and view data from the analyzer. The data may consist of SO2 concentration, or may be diagnostic data, such as the sample flow or PMT output.

The M100AH comes pre-programmed with a set of useful Data Channels for logging SO2 concentration and predictive diagnostic data. The default Data Channels can be used as they are, or they can be changed by the user to fit a specific application. They can also be deleted to make room for custom user-programmed Data Channels.

5-12

Page 56
Image 56
Teledyne 100AH Set-Up Mode, Configuration Information CFG, Automatic Calibration AutoCal, Data Acquisition System DAS