Teledyne 100AH instruction manual Negative Concentration Display, Excessive Noise, 10-25

Models: 100AH

1 172
Download 172 pages 44.73 Kb
Page 133
Image 133

Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

10.2.4 Negative Concentration Display

1.Mis-calibration. The 'zero' gas that was used to zero the M100AH contained some SO2 gas - that is it had more SO2 gas than that of the sample air. May also be caused by doing a zero calibration using ambient air.

2.Broken PMT temperature control circuit, causing high zero offset. Check PMT temperature which should be 7 ± 1°C.

3.Aging UV filter causing high stray light. Check STR. LGT reading of TEST function. It should be less than 100 PPM.

4.Check for light leak.

10.2.5 Excessive Noise

Common reasons for excessive noise are:

1.Leak in pneumatic system.

2.Light leak - check the sensor module with strong light.

3.HVPS noisy - see HVPS test procedure. See Section 10.3.9.

4.Defective electronic components on preamp board. - use optic test and electric test to check electronics, optics and observe noise.

5.Contamination of reaction cell and optical system - This can be wet air or impurities. This can be detected by high PMT readings with zero air as sample gas. Clean reaction cell as described in Section 9.5.

6.Broken PMT temperature control circuit. Check PMT TEMP - TEST function.

7.Mis-calibration. Check SLOPES in TEST function.

8.Too low UV lamp output. Should be higher than 600 mV. Replace the lamp if it is lower than this value.

9.High stray lights - UV filter is aging.

10.Reaction cell temperature is not stable.

10-25

Page 133
Image 133
Teledyne 100AH instruction manual Negative Concentration Display, Excessive Noise, 10-25