Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

Example 3: To perform zero-span calibration check once per day at 10:30 PM and zero calibration adjust once per week starting at 11:30 PM, 5/20/97.

1.Select any one of SEQx and setup as example 1 above.

2.Select any other SEQx and program as follows. Always avoid setting two or more sequences at the same time of the day.

1)MODE: ZERO

2)TIMER ENABLE: ON

3)STARTING DATE: 5/20/97

4)STARTING TIME: 23:30

5)DELTA DAYS: 7

6)DELTA TIME: 00:00

7)DURATION: 15.0 MINUTES

8)CALIBRATE: ON

6-5

Page 69