Teledyne 100AH instruction manual Enter Expected Span Gas Concentration Procedure, Cal-Conc

Models: 100AH

1 172
Download 172 pages 44.73 Kb
Page 92
Image 92

Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

Enter the expected SO2 span gas concentration:

Table 8-3: Enter Expected Span Gas Concentration Procedure

Step Number

Action

Comment

1.

Press

This key sequence causes the M100AH to prompt for the

 

CAL-CONC

expected SO2 concentration. Enter the SO2 span

 

 

concentration value by pressing the key under each digit

 

 

until the expected value is set. This menu can also be entered

 

 

from CALS.

 

 

 

2.

Press ENTR

ENTR stores the expected SO2 span value.

4.

Press EXIT

Returns instrument to SAMPLE mode.

 

 

 

Table 8-4: Manual Span Calibration Procedure - Span Gas Through Sample Port

Step Number

Action

Comment

1.

Press CAL

The M100AH enters the calibrate mode. External span gas

 

 

should be fed to the sample port. If AutoRange is enabled,

 

 

then select LO or HI range and press ENTR.

 

 

 

2.

Wait 10 min

Wait for reading to stabilize at span value.

 

 

 

3.

Press SPAN

If you change your mind after pressing SPAN, you can still

 

 

press EXIT here without spanning the instrument.

 

 

 

4.

Press ENTR

Pressing ENTR actually changes the calibration equations

 

 

and causes the instrument to read the SO2 span

 

 

concentrations.

 

 

 

5.

Press EXIT

M100AH returns to sampling. Immediately after calibration,

 

 

data is not added to the DAS averages.

 

 

 