Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

1. Calibration Sequence Mode

Each sequence can generate any one of 7 different combinations of ZERO, LO, or HI span point. Press SETUP-ACAL, and scroll up or down to select the desired sequence number (SEQx). Press MODE and scroll up or down by pressing PREV or NEXT. Select one of the combination shown below and press ENTR.

Combinations:

1)DISABLED; will disable corresponding SEQx setup.

2)ZERO

3)ZERO-LO

4)ZERO-HI

5)ZERO-LO-HI

6)LO

7)HI

8)LO-HI

2.Setup Calibration Timer:

Press SETUP-ACAL-SET to setup or edit the automatic calibration timer.

Following table summarizes the setup procedures:

Table 6-2: Setup Automatic Zero/Span Calibration

 

Default

Description

Timer Enable

ON

Enable or disable automatic calibration timer.

 

 

 

Starting Date

01-JAN-95

MM:DD:YY

 

 

 

Starting Time

00:00

HH:MM, 0 - 23 hours and 0 - 59 minutes

 

 

 

Delta Days

1

Delay days between each calibration

 

 

(0 - 365 days)

 

 

 

Delta Time

00:00

HH:MM, 0 - 23 hours and 0 - 59 minutes

 

 

 

Duration

15.0 minutes

1 - 60.0 minutes

 

 

 

Calibrate

OFF

on/off. If ON is selected, it will adjust the

 

 

calibration.

 

 

 

The Timer Enable can be set to “OFF” to disable the automatic calibration timer while the remote RS-232 calibration of specific sequence can be initiated.

6-3

Page 67
Image 67
Teledyne 100AH instruction manual Setup Automatic Zero/Span Calibration, Default Description, Mmddyy, Off