Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D
1. Calibration Sequence Mode
Each sequence can generate any one of 7 different combinations of ZERO, LO, or HI span point.
Press SETUP-ACAL, and scroll up or down to select the desired sequence number (SEQx).
Press MODE and scroll up or down by pressing PREV or NEXT. Select one of the combination
shown below and press ENTR.
Combinations:
1) DISABLED; will disable corresponding SEQx setup.
2) ZERO
3) ZERO-LO
4) ZERO-HI
5) ZERO-LO-HI
6) LO
7) HI
8) LO-HI
2. Setup Calibration Timer:
Press SETUP-ACAL-SET to setup or edit the automatic calibration timer.
Following table summarizes the setup procedures:
Table 6-2: Setup Automatic Zero/Span Calibration
Default Description
Timer Enable ON Enable or disable automatic calibration timer.
Starting Date 01-JAN-95 MM:DD:YY
Starting Time 00:00 HH:MM, 0 - 23 hours and 0 - 59 minutes
Delta Days 1 Delay days between each calibration
(0 - 365 days)
Delta Time 00:00 HH:MM, 0 - 23 hours and 0 - 59 minutes
Duration 15.0 minutes 1 - 60.0 minutes
Calibrate OFF on/off. If ON is selected, it will adjust the
calibration.
The Timer Enable can be set to “OFF” to disable the automatic calibration timer while the
remote RS-232 calibration of specific sequence can be initiated.
6-3