Table 10-4: Diagnostic Mode - Signal I/O (Continued)

No.

Signal

Control

Description

19.

ST_TEMP_ALARM

YES

Status Bit - Temperature alarm

 

 

 

Logic High = Reaction cell, PMT, Box temps out of

 

 

 

spec

 

 

 

Logic Low = Temps within spec

 

 

 

 

20.

ST_DIAG_MODE

YES

Status Bit - In Diagnostic mode

 

 

 

Logic High = M100AH in Diagnostic mode

 

 

 

Logic Low = Not in Diag mode

 

 

 

 

21.

ST_POWER_OK

YES

Status Bit - Power OK

 

 

 

Logic High = Instrument power is on

 

 

 

Logic Low = Instrument power is off

 

 

 

 

22.

ST_SYSTEM_OK

YES

Status Bit - System OK

 

 

 

Logic High = No instrument warning present

 

 

 

Logic Low = 1 or more alarm present

 

 

 

 

23.

ST_HVPS_ALARM

YES

Status Bit - HVPS alarm

 

 

 

Logic High = HVPS out of spec

 

 

 

Logic Low = HVPS within spec

 

 

 

 

24.

PMT_SIGNAL

NO

Current PMT voltage. Same as PMT voltage in

 

 

 

TEST menu. Bi-polar, typically in 0-5000 mV

 

 

 

range. A constant value of 5000 mV indicates off

 

 

 

scale.

25.

RCELL_TEMP

NO

Reaction Cell temperature. 3500 mV for 50°C.

 

 

 

 

26.

BOX_TEMP

NO

Box Temperature. Typically 1800 mV for 25°C.

 

 

 

 

27.

PMT_TEMP

NO

PMT cold block temperature. Typically 1800 mV

 

 

 

for 7°C.

 

 

 

 

28.

DCPS_VOLTAG

NO

DC power supply composite voltage output.

 

 

 

Typically 2500 mV.

 

 

 

 

29.

SAMPLE_PRES

NO

Sample inlet pressure in mV. Typical sea level

 

 

 

value = 3750 mV for 29.9" HG-A.

30

VACCUM_PRESS

NO

Reaction cell pressure in mV. Typical value = 750

 

 

 

mV for 6” HG.