Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

7.7 DAS Commands and Reports

RS-232 Commands

In addition to accessing the data acquisition system and the stored data from the instrument front panel, you can also access the data acquisition and the stored data from the RS-232 interface.

There are two RS-232 commands, listed in the table below.

Table 7-12: RS-232 DAS Commands

 

RS-232 DAS Commands

Command

 

Description

D [id] PRINT [“name”]

 

Prints specified data channel’s

 

 

properties

 

 

 

D [id] REPORT “name”

 

Prints the data for the specified data

[RECORDS=number]

 

channel

[COMPACTVERBOSE]

 

 

 

 

 

In all of the commands, brackets ([ ]) denote optional parameters. The ID parameter is the instrument ID, useful when the multi-drop protocol is being used. The NAME parameter is the data channel’s name. It must be enclosed in quotes (i.e. D PRINT “CONC”).

The RECORDS parameter of the REPORT command indicates how many records from the most recent record and prior to print. If the RECORDS parameter is not specified, all of the records are printed. The COMPACT and VERBOSE parameters of the REPORT command specify the report format.

RS-232 Reports

There are two basic kinds of RS-232 reports: data channel summary report, and data reports.

Data Channel Summary Format

When you press the PRNT button in the data channel edit menu, a report like the following is printed on the RS-232 channel:

7-14

Page 84
Image 84
Teledyne 100AH instruction manual DAS Commands and Reports, 12 RS-232 DAS Commands, Compactverbose