Teledyne 100AH instruction manual Manual Zero Calibration Procedure Z/S Valves

Models: 100AH

1 172
Download 172 pages 44.73 Kb
Page 93
Image 93

Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

8.2Manual Zero/Span Check or Calibration with Zero/Span Valves Option

The Zero/Span valve option can be operated from the front panel keyboard. In the Zero/Span valve option the zero and span gas comes into the valves through ports on the rear panel of the instrument.

Table 8-5: Manual Zero Calibration Procedure - Z/S Valves

Step Number

Action

Comment

1.

Press CALZ

The analyzer enters the zero calibrate mode. This switches

 

 

the zero valve to allow zero gas to come in through the zero

 

 

gas inlet port on the rear panel. If AutoRange is enabled,

 

 

then select LO or HI range and press ENTR.

 

 

 

2.

Wait 10 min

Wait for reading to stabilize at zero value.

 

 

 

3.

Press ZERO

If you change your mind after pressing ZERO, you can still

 

 

press EXIT here without zeroing the instrument.

 

 

 

4.

Press ENTR

Pressing ENTR actually changes the calculation equations,

 

 

forcing the reading to zero.

5.

Press EXIT

M100AH returns to sample mode. Immediately after

 

 

calibration, readings do not go into the DAS averages.

Table 8-6: Manual Span Calibration Procedure - Z/S Valves

Step Number

Action

Comment

1.

Press CALS

The M100AH enters the calibrate mode from sample mode.

 

 

This operates the sample/cal and zero/span valves to allow

 

 

span gas to come in through the cal gas inlet port or the rear

 

 

panel. If AutoRange is enabled, then select LO or HI range

 

 

and press ENTR.

 

 

 

2.

Wait 10 min

Wait for reading to stabilize at span value.

 

 

 

3.

Press SPAN

If you change your mind after pressing SPAN, you can still

 

 

press EXIT here without spanning the instrument.

 

 

 

4.

Press ENTR

Pressing ENTR actually changes the calculation equations.

 

 

 

5.

Press EXIT

M100AH returns to sampling. After calibration, data is not

 

 

added to the DAS averages during HOLDOFF period.

 

 

 

8-5

Page 93
Image 93
Teledyne 100AH Manual Zero Calibration Procedure Z/S Valves, Manual Span Calibration Procedure Z/S Valves