Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

2.3Initial Operation

1.Turn on the instrument power.

2.The display should immediately light, displaying the instrument type (M100AH) and the CPU memory configuration. If you are unfamiliar with the M100AH, we recommend that you read the overview Section 4 before proceeding. A diagram of the software menu trees is in Figure 5-1and Figure 5-2.

3.The M100AH requires about 30 minutes for all internal components to come up to temperature.

4.While waiting for instrument temperatures to come up, you can check for correct operation by using some of the M100AH's diagnostic and test features.

5.Examine the TEST functions by comparing the values listed in Table 2-1to those in the display. Remember that as the instrument warms up the values may not have reached their final values yet. If you would like to know more about the meaning and utility of each TEST function refer to Table 10-1. Table 2-1also contains the list of options. Section 6 covers setting up the options.

6.When the instrument is warmed up, re-check the TEST functions against Table 2-1.All of the readings should compare closely with those in the Table. If they do not, see

Section 10.1.1.

2-7