Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D
Table 2-1: Final Test and Calibration Values
Test Values Observed
Value Units Nominal Range Reference Section
RANGE PPM 10 - 5000 5.3.4
STABIL PPM 0.01 - 0.05 5.2.1, 10.1.1,
10.2.5, Table 10-1
INLET
SAMPLE
PRESS
in-Hg-
Abs
25 - 35 10.1.1, 10.3.5,
Table 10-1
VACUUM
PRESS
in-Hg-
Abs
4 - 10 10.3.5, Table 10-1
SAMP FL CC / MIN 650 ± 60 10.2.2, 10.3.5,
Table 10-1
PMT mV 0 - 5000 10.4.1
UV LAMP mV 2000 - 4000
typical
10.4.2
STR. LGT PPM <100.0 Table 10-1
DARK PMT mV <200 Table 10-1
DARK LAMP mV <200 Table 10-1
SLOPE 1.0 ± 0.3 8.7
OFFSET mV <200 8.7
HVPS V 450 - 900 constant 10.3.9
DCPS mV
2500 ± 200 10.3.4
RCELL TEMP oC 50 ± 1 10.3.6
BOX TEMP oC 8 - 50 10.3.3.1
PMT TEMP oC 7 ± 1 10.3.8
Electric Test & Optic Test
Electric Test
PMT Volts mV 2000 ± 100 10.1.3.2
SO2 Conc PPM 2000 ± 100 10.1.3.2
Optic Test
PMT Volts mV 200 ± 20 10.1.3.3
SO2 Conc PPM 200 ± 20 10.1.3.3
2-12