Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

Table 2-1: Final Test and Calibration Values

Test Values

Observed

Units

Nominal Range

Reference Section

Value

 

 

 

 

RANGE

 

PPM

10 - 5000

5.3.4

 

 

 

 

 

STABIL

 

PPM

0.01 - 0.05

5.2.1, 10.1.1,

 

 

 

 

10.2.5, Table 10-1

 

 

 

 

 

INLET

 

in-Hg-

25 - 35

10.1.1, 10.3.5,

SAMPLE

 

Abs

 

Table 10-1

PRESS

 

 

 

 

 

 

 

 

 

VACUUM

 

in-Hg-

4 - 10

10.3.5, Table 10-1

PRESS

 

Abs

 

 

 

 

 

 

 

SAMP FL

 

CC / MIN

650 ± 60

10.2.2, 10.3.5,

 

 

 

 

Table 10-1

PMT

 

mV

0 - 5000

10.4.1

 

 

 

 

 

UV LAMP

 

mV

2000 - 4000

10.4.2

 

 

 

typical

 

 

 

 

 

 

STR. LGT

 

PPM

<100.0

Table 10-1

 

 

 

 

 

DARK PMT

 

mV

<200

Table 10-1

 

 

 

 

 

DARK LAMP

 

mV

<200

Table 10-1

 

 

 

 

 

SLOPE

 

 

1.0 ± 0.3

8.7

 

 

 

 

 

OFFSET

 

mV

<200

8.7

 

 

 

 

 

HVPS

 

V

450 - 900 constant

10.3.9

 

 

 

 

 

DCPS

 

mV

2500 ± 200

10.3.4

 

 

 

 

 

RCELL TEMP

 

oC

50 ± 1

10.3.6

BOX TEMP

 

oC

8 - 50

10.3.3.1

PMT TEMP

 

oC

7 ± 1

10.3.8

Electric Test & Optic Test

Electric Test

PMT Volts

 

mV

 

2000 ± 100

10.1.3.2

 

 

 

 

 

 

 

SO2

Conc

 

PPM

 

2000 ± 100

10.1.3.2

 

 

 

 

Optic Test

 

PMT Volts

 

mV

 

200 ± 20

10.1.3.3

 

 

 

 

 

 

 

SO2

Conc

 

PPM

 

200 ± 20

10.1.3.3

2-12

Page 24
Image 24
Teledyne 100AH Final Test and Calibration Values, Test Values Observed Units Nominal Range Reference Section, Optic Test