Teledyne 100AH instruction manual RS-232 Warning Messages

Models: 100AH

1 172
Download 172 pages 44.73 Kb
Page 79
Image 79

Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

The T command lists TEST messages. Examples of the T command are:

T LIST

Lists test messages currently enabled with T SET

T LIST ALL

Lists all test messages

T RCELLTEMP

Prints the temperature of the reaction cell

T SO2CONC

Prints SO2 concentration message

T LAMPRATIO

Prints Lamp Ratio

7.4 WARNING Commands and Messages

Table 7-8: RS-232 Warning Messages

Name

Message

Description

WSYSRES

SYSTEM RESET

Analyzer was reset/powered on

 

 

 

WRAMINIT

RAM INITIALIZED

RAM was erased

 

 

 

WSAMPFLOW

SAMPLE FLOW WARNING

Sample flow out of spec.

 

 

 

WSAMPPRESS

SAMPLE PRESSURE

Sample pressure below 15” Hg or

 

WARNING

above 35” Hg

WVACPRESS

VACUUM PRESSURE

Vacuum pressure below 1 in-Hg or

 

WARNING

above 10 in-Hg.

WPMT

PMT DET WARNING

PMT output above 4995 mV.

 

 

 

WUVLAMP

UV LAMP WARNING

UV lamp output is below 600 mV or

 

 

above 4995 mV

 

 

 

WDARKCAL

DARK CAL WARNING

Dark PMT OR Dark UV lamp is

 

 

above 400 mV

 

 

 

WPMTTEMP

PMT TEMP WARNING

PMT temperature too high/low

 

 

 

WRCELLTEMP

RCELL TEMP WARNING

Reaction cell temp. out of spec.

 

 

 

WBOXTEMP

BOX TEMP WARNING

Box temperature too high/low

 

 

 

WDYNZERO

CANNOT DYN ZERO

Dynamic zero cal. out of spec.

 

 

 

WDYNSPAN

CANNOT DYN SPAN

Dynamic span cal. out of spec.

 

 

 

WHVPS

HVPS WARNING

HVPS too high/low

 

 

 

WVFDET

V/F NOT INSTALLED

A/D board not installed or broken

 

 

 

WDCPS

DCPS WARNING

DC power supply output below

 

 

2300 mV or above 2700 mV

 

 

 

7-9