Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

7.5 CALIBRATION Commands and Messages

Table 7-9: RS-232 Calibration Messages

Message

Description

START ZERO CALIBRATION

Beginning IZS zero calibration

 

 

FINISH ZERO CALIBRATION, SO21=xxxxx PPM

Finished IZS zero calibration

START SPAN CALIBRATION

Beginning IZS span calibration

 

 

FINISH SPAN CALIBRATION, SO21=xxxxx PPM

Finished IZS span calibration

START MULTI-POINT CALIBRATION

Beginning multi-point calibration

 

 

FINISH MULTI-POINT CALIBRATION

Finished multi-point calibration

 

 

1Depends on which units are currently selected.

 

Whenever the analyzer starts or finishes an Zero/Span calibration, it issues a status report to the RS-232 interface. If the RS-232 interface is in the normal mode, these reports will be sent. Otherwise, they will be discarded. The format of these messages is:

C DDD:HH:MM IIII CALIBRATION STATUS MESSAGE

An example of an actual sequence of calibration status messages is:

C DDD:HH:MM IIII START MULTI-POINT CALIBRATION

C DDD:HH:MM IIII FINISH MULTI-POINT CALIBRATION

There are several methods of both checking the calibration and calibrating the M100AH that are discussed in Section 8. The C command executes a calibration command, which may be one of the following:

7-11