Teledyne API Model 100AH SO2 Analyzer Instruction Manual, 02417, Rev. D

Table 10-1: Test Functions (Continued)

Test Function

Factory Set-Up

Comment

DCPS

2500 ± 200 mV

DCPS is a composite of the +5 and ± 15 VDC supplies. It has

 

 

been arbitrarily set at 2500 ± 200 mV. If it is not in this range

 

 

one of the voltages in the supply is not working. Check the

 

 

procedures for diagnosing the Power Supply Module.

 

 

 

RCELL TEMP

50.0 ± 1°C

The reaction cell temperature is controlled to 50°C ± 1°C by

 

 

the computer. It should only read other values when the

 

 

instrument is warming up. If the value is outside the

 

 

acceptable range, go to the procedure for diagnosing the

 

 

Reaction cell temp supply. The alarm limits are less than 45°C

 

 

and greater than 55°C.

 

 

 

BOX TEMP

8 - 48°C

The Box Temp is read from a thermistor on the Status/Temp

 

 

board (01086). It should usually read about 5°C above room

 

 

temp. The M100AH is designed to operate from 5 to 40°C

 

 

ambient. Therefore the box temperature should be in the range

 

 

of about 10 to 50°C. Temperatures outside this range will

 

 

cause premature failures of components, and poor data

 

 

quality. Warning limits are < 8°C and > 52°C.

 

 

 

PMT TEMP

7.0 ± 1°C

The PMT detector is very temperature sensitive. The PMT

 

 

temperature should always be near 7°C, except at power-up.

 

 

Temperatures more than ± 1°C from the set point indicate

 

 

problems with the cooler circuit. See Section 10.3.9 for PMT

 

 

cooler diagnostic and troubleshooting. Warning limits are <

 

 

2°C and > 12°C.

 

 

 

TIME

 

This is the time of day clock readout. It is used to time the

 

 

AutoCal cycles. The speed of the clock can be adjusted by the

 

 

CLOCK_ADJ variable in the VARS menu. The clock can be

 

 

set via SETUP-CLOCK-TIME from the front panel.

 

 

 