Agilent Technologies N9340A manual Error Message List, Error Messages

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 154
Image 154

7 Error Messages

Error Message List

 

N O T E

A symbol [F] at the beginning of an error message

 

indicates this is an error caused by manual front

 

 

 

 

panel operation. A symbol [R] at the beginning of

 

 

an error message indicates this is an error caused

 

 

by remote control mode.

 

 

 

 

 

 

 

 

Error Code Error Message

Description

 

0

No error

 

 

 

–410

Query INTERRUPTED

Indicates that a condition causing

 

 

 

an INTERRUPTED query

 

 

 

occurred (see IEEE 488.2,

 

 

 

6.3.2.7)

 

 

–350

Query overflow

Indicates the SCPI remote

 

 

 

interface error queue

 

 

 

overflowed.

 

–321

Out of memory

Indicates an internal operation

 

 

 

needed more memory than that

 

 

 

was available.

 

–224

Illegal

Indicates you sent a parameter

 

 

parameter value

for this command that is NOT

 

 

 

allowed.

 

–223

Too much data

Indicates a legal program data

 

 

 

element of block, expression or

 

 

 

string type was received that

 

 

 

contained more data than the

 

 

 

device could handle due to

 

 

 

related device-specific

 

 

 

requirements or memory.

 

–222

Data out of

Indicates a legal data was parsed

 

 

range

but could not be executed

 

 

 

because of the interpreted value

 

 

 

was outside the legal range

 

 

 

defined by the analyzer. The

 

 

 

displayed results may be clipped.

144

N9340A User’s Guide

Page 154
Image 154
Agilent Technologies N9340A manual Error Message List, Error Messages