Agilent Technologies N9340A manual Channel Power Subsection, Int BW, Channel Span, Power Density

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 132
Image 132

6 SCPI Command Reference

Channel Power Subsection

Center Freq

:MEASure:CHPower:CENTer <freq>

:MEASure:CHPower:CENTer?

Sets the center frequency of the display.

Int BW

:MEASure:CHPower:IBW <freq>

:MEASure:CHPower:IBW?

Specifies the integration bandwidth ranging from 100 Hz to 3 GHz to calculate the power.

Channel Span

:MEASure:CHPower:SPAN <freq>

:MEASure:CHPower:SPAN?

Sets the analyzer span for the channel power measurement. Be sure the span is set between 1 and 10 times the integration bandwidth.

There are also some remote commands which are irrelative to the front- panel keys.

Channel Power and Power Density

:MEASure:CHPower?

Returns channel power and power density.

Channel Power

:MEASure:CHPower:CHPower?

Returns the channel power.

Power Density

:MEASure:CHPower:DENSity?

Returns power density.

122

N9340A User’s Guide

Page 132
Image 132
Agilent Technologies N9340A Channel Power Subsection, Int BW, Channel Span, Channel Power and Power Density, Center Freq