Agilent Technologies N9340A Amplitude Step, Amplitude Offset, TGENeratorAMPLitudeOFFSet ampt

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 146
Image 146

6 SCPI Command Reference

Amplitude Step

:TGENerator:AMPLitude:STEP <ampt> :TGENerator:AMPLitude:STEP?

This command sets the amplitude step size of the tracking generator output. This command is valid after tracking generator is enabled.

*RST State:

1 dB

Range:

1 to 10 dB

Min. Increment

1 dB

Available unit:

dB

Key Access:

{Amptd Step}

 

 

Amplitude Offset

:TGENerator:AMPLitude:OFFSet <ampt>

This command sets the amplitude offset of the tracking generator output. This command is valid after tracking generator is enabled.

*RST State:

0.00 dB

Range:

–88.40 to 327.60 dB

Min. Increment

0.01 dB

Available unit:

dB

Key Access:

{Amptd Offset}

 

 

136