Fast

 

Activates the fast sweep mode.

 

Fast sweep mode provides a fast measurement

 

function that decreases the sweep time. However,

 

this mode brings a decrease of amplitude accuracy.

 

Fast sweep is valid only when span is no less than 50 MHz.

N O T E

Measurement data is not ganruanteed in fast sweep mode.

 

 

 

 

Key access: [BW/SWP] > {Sweep} > {Sweep Mode}

 

Single Sweep

 

When the analyzer is in continuous sweep mode

 

and not in a measurement mode ([MEAS] > Measure

 

Off), pressing [BW/SWP] > {Single Sweep} convert the

 

continuous sweep to single sweep and executes a

 

sweep after the trigger condition is met. If the

 

analyzer is already in single sweep, pressing

 

[BW/SWP] > {Single Sweep} executes a new sweep as

 

soon as the trigger condition is met.

 

If Average is on ([TRACE] > {More (1 of 2)} > {Average}

 

> On), pressing [BW/SWP] > {Single Sweep} halts

 

averaging of the current value, and continue

 

averaging from the current count number as soon

 

as a single sweep function triggers.

 

Key access: [BW/SWP] > {Sweep}