5Key Reference
•Fast
| Activates the fast sweep mode. | |
| Fast sweep mode provides a fast measurement | |
| function that decreases the sweep time. However, | |
| this mode brings a decrease of amplitude accuracy. | |
| Fast sweep is valid only when span is no less than 50 MHz. | |
N O T E | ||
Measurement data is not ganruanteed in fast sweep mode. | ||
| ||
|
| |
| Key access: [BW/SWP] > {Sweep} > {Sweep Mode} | |
| Single Sweep | |
| When the analyzer is in continuous sweep mode | |
| and not in a measurement mode ([MEAS] > Measure | |
| Off), pressing [BW/SWP] > {Single Sweep} convert the | |
| continuous sweep to single sweep and executes a | |
| sweep after the trigger condition is met. If the | |
| analyzer is already in single sweep, pressing | |
| [BW/SWP] > {Single Sweep} executes a new sweep as | |
| soon as the trigger condition is met. | |
| If Average is on ([TRACE] > {More (1 of 2)} > {Average} | |
| > On), pressing [BW/SWP] > {Single Sweep} halts | |
| averaging of the current value, and continue | |
| averaging from the current count number as soon | |
| as a single sweep function triggers. | |
| Key access: [BW/SWP] > {Sweep} |
78 | N9340A User’s Guide |