5 Key Reference

Ref Offset

 

Adds an offset value which ranges from - 327.6 dB

 

to +327.6 dB to the displayed reference level.

 

Reference-level offsets are only entered by using the numeric

N O T E

keypad. Entering an offset does not affect the trace or the

 

 

attenuation value.

 

 

 

Reference- level offsets are used when gain or loss

 

occurs between a device under test and the

 

analyzer input. Thus, the signal level measured by

 

the analyzer may be referred to as the level at the

 

input of an external amplitude conversion device.

 

When an amplitude offset is entered, its value

 

appears in the lower left corner of the display.

 

Key access: [AMPTD]

Y Axis Units

 

Sets the amplitude units.

 

Amplitude units are maintained for both logarithmic and linear

N O T E

modes depending on the Scale Type setting.

 

 

 

 

Key access: [AMPTD] > {Y Axis Units}

72

N9340A User’s Guide