Getting Started | 2 |
Viewing a Signal
1Use a signal generator to generate a continuous- wave signal (1 GHz, 0 dBm).
2Press [SYS] > {More (1 0f 3)} > {Preset} and select
Default to toggle the preset setting to factory- defined status.
3Press the green [Preset] key to restore the analyzer to its factory- defined setting.
4Connect the generator’s RF OUT connector to analyzer’s RF IN connector on the top panel.
5Press [FREQ] > {Center Freq} > 1> {GHz} to set the center frequency to 1 GHz.
6Press [SPAN] > 5 > {MHz} to set the frequency span to 5 MHz.
7Press [MARKER] > {Peak Search} > {Peak} to place a marker (labeled 1) at the highest peak (1 GHz) on the display.
The values of amplitude and frequency of the Marker appear both in the function block and in the upper- right corner of the screen.
Use the knob, the arrow keys, or the softkeys in the Peak Search menu to move the marker and read out the value of both frequency and amplitude displayed on the screen.
Figure 1 View a signal (1 GHz, 0 dBm)
N9340A User’s Guide | 21 |