Agilent Technologies N9340A Marker Trace, All Off, Noise, power density, Clear write, Max hold

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 97
Image 97
• Noise

Key Reference 5

 

• Noise

 

Activates a noise readout mode for evaluating

 

power density.

N O T E

Noise mode operates only on noise peaks and not supported

under Marker Delta condition.

 

 

Key access: [MARKER] > {More (1 of 2)} > {Mode}

Marker Trace

Activates a marker on the trace if there are no markers turned on. If a marker is currently active, press Marker Trace until 1, 2, 3, or 4 is underlined. The active marker will be moved to the selected trace.

Selecting Auto will move the marker to the automatically selected trace. The selection order is to look for the lowest numbered trace in the following order of trace modes:

 

 

Clear write

Max hold

Min. hold

View

If there is no trace in any of these modes, it selects trace 1 as default.

Key access: [MARKER] > {Marker Trace}

All Off

Turns off all of the markers, including markers used for signal track and demodulation. Marker annotation is also removed.

Key access: [MARKER] > {All Off}

N9340A User’s Guide

87

Page 97
Image 97
Agilent Technologies N9340A Marker Trace, All Off, Noise, Activates a noise readout mode for evaluating, power density