Minimum Hold

Key Reference 5

Minimum Hold

 

Maintains the minimum level for each trace point

 

of the selected trace, and updates each trace point

 

if a new minimum level is detected in successive

 

sweeps.

 

Changing the vertical scale (Amplitude, Scale Type, Log or

N O T E

Line), or pressing PRESET, or turning averaging on (TRACE,

 

 

Average (On)) or switching widow in Zone Span restarts the

 

held trace.

Key access: [TRACE]

View

Holds and displays the amplitude data of the selected trace. The trace register is not updated as the analyzer sweeps. If a trace is deactivated by pressing Blank, the stored trace data can be retrieved by pressing View.

Key access: [TRACE]

Blank

Stores the amplitude data for the selected trace and removes it from the display. The selected trace register will not be updated as the analyzer sweeps. This function is activated for traces 2, 3 and 4 at power- on and Factory Preset.

Key access: [TRACE]

Detector

Selects a specific detector, or uses the system to pick the appropriate detector for a particular measurement.

When discussing detectors, it is important to understand the concept of a trace “bucket”. For every trace point displayed, there is a finite time during which the data for that point is collected. The analyzer has the ability to look at all of the

N9340A User’s Guide

91